GEOLOGICAL SURVEY CIRCULAR 544
Determination of Gold in
Geologic Materials by
Solvent Extraction and
Atomic-Absorption
Spectrometry
Determination of Gold in
Geologic Materials by.
Solvent Extraction and
Atomic-Absorption
Spectrometry
By Claude Huffman, Jr., J. D. Mensik, and
L. B. Riley
Mセ@
GEOLOGICAL SURVEY CIRCULAR 544
Washington 1967
United States Department of the Interior
STEWART L. UDALL, Secretary
Geological Survey
William T. Pecora, Director
First printing 1967
Second printing 1 968
Free on application to the U.S. Geological Survey, Washington, D.C. 20242
Abstract ——————————-Introduction
—————————-Reagents——————————-
Apparatus——————————
Cyanide
atomic-absorption procedure ——
Sample preparation——————–Standardization
———- ————–
CONTENTS
Page
1
1
1
2
3
3
4
Fire-assay atomic-absorption procedure —
Results ———————————
Sample dissolution, cyanide procedure—–Extraction
of gold with methyl isobutyl
ketone——————————–
Conclusions – – – – – – – – – – – – – – – – – – – – – – — – – – -References——————————
ILLUSTRATIONS
Page
4
5
5
5
6
6
Page
Figure 1. Water bath equipped for aeration and constant temperature—————————– 2
2. Constant-temperature water bath with magnetic stirrer——————————– 3
TABLES
Page
Table 1. Comparison of gold values obtained by atomic-absorption methods and by the
conventional gravimetric fire-assay method —————————————- 5
2. Comparison of gold values obtained by fire-assay atomic-absorption and
cyanide atomic-absorption methods———————————————— ·6
III
Determination of Gold in Geologic Materials by
Solvent Extracti n and Atomic-Absorption Spectrometry
Huffman, Jr., J. D. Mens’ik, and L. B. Riley
ABSTRACT
The two methods presented for the deter ination of traces of
gold in geologic materials are the cyani e atomic-absorption
method and the fire-assay atomic-absorption ethod. In the cyanide
method gold is 1 each e d with a sodium c y n ide solution. The
monovalent gold is then oxidized to the trival nt state and concentrated
by extracting into methyl isobutyl keto e prior to estimation
by atomic absorption. In the fire-assay atomi -absorption method,
the gold-silver bead obtained from fire assay· s dissolved in nitric
and hydrochloric acids. Gold is then conce trated by extracting
into methyl isobutyl ketone prior to determ ·nation by atomic absorption.
By either method concentrations a low as 50 parts per
billion of gold can be determined in a 15-gra sample.
INTRODUCTION
The U.S. Geological Survey heav metals program
has instituted reconnaissance studie requiring rapid,
precise, and sensitive methods for de ermining various
elements in geologic materials. The determination of
gold for this program has required ethods more sensitive
than the conventional fire-ass y method. Lakin
and Nakagawa (1965) described a sp ctrophotometric
method for determining traces o gold in geologic
materials; they found, however (oral commun., 1966),
that thallium and iron interfere. Wals ( 1955) has shown
that atomic-absorption spectrometry is extremely well
suited for direct determination of low concentrations of
certain elements in solution, but until recently no work
seems to have been undertaken forth determination of
gold. Olson (1965) and Simmons (19 5) used a cyanide
solution to leach the gold from silice us carbonate ores
prior to estimation by atomic absorpt on. Tindall ( 1965)
described an atomic-absorption met od for the determination
of gold in various types of amples dealt with
in the copper mining industry. He u ed an aqua regia
digestion followed by an organic ext action toconcentrate
the gold into a small volume Strelow, Feast,
Mathews, andVanZyl(1966)applied natomic-absorption
method to the determination o gold in cyanide
waste solutions. All these atomic-a sorption methods
have been used primarily for deter:nining gold in the
parts-per-million range.
This paper describes two rapid 。ョ、ーイ・」ゥウLセエッュᆳ
absorption methods for gold determination currently
used in the U.S. Geological Survey’s Denver laboratory:
the cyanide atomic-absorption method and エィセ@ fire-assay
atomic-absorption method. In developing these
methods, sensitivity was stressed to enable g?.ochemical
prospecting for gold at theparts-per-billionlevel.
Both methods can detect concentrations as low as 50
ppb (parts per billion) or 0.0015 ounce per ton of gold
in a 15-gram sample.
In the cyanide method the sample is roasted and then
treated with a sodium cyanide solution to dis·solve the
gold. The cyanide solution is acidified with tydrobromic
acid and bromine water to oxidize the m’)novalent
gold to the trivalent state. The gold is then con7.entrated
by extracting into a measured volume of methyl isobutyl
ketone (MIBK). The gold concentration is determined by
an atomic -absorption spectrophotometer.
In the classical fire-assay method, the sil·rer ofthe
gold-silver bead obtained by cupellation is dissolved
and the gold residue is determined gravim<?.trically.
The sensitivity of this method is approximately 0.02
troy ounce per ton (600 ppb) on a 15-g (half assay ton)
sample. In the proposed fire-assay atomic-absorption
method the gold-silver bead from fire assay is dissolved
in nitric and hydrochloric acids anc then extracted
into a small volume of organic solve·1t (MIBK)
prior to estimation by atomic absorption. This procedure
thus increases the sensitivity by about a factor
of 10 when a 15-g sample is used. The sensitivity can
be further increased by combining beads obtained from
two fire-assay fusions of 30 g each of the same sample.
With this technique about 15 ppb can be determined.
This is about the limit imposed by reagent blanks.
REAGENTS
Standard gold stock solution, 400 ppm (parts per million)
gold. Dissolve 0.200 g of gold metal in a covered
beaker with 30 milliliters of aqua regia on a steam
bath, remove cover, and evaporate to dryress. Heat
on a hotplate until fumes from HN03 are no longer
present. Add 250 ml of ( 1:1) HCl to the beaker and
digest on the steam bath to dissolve the gold salts.
Transfer the solution to a 500-ml volumetric flask
and dilute to volume with distilled water.
Sodium cyanide solution, 0.25 percent. Dissolve 5 g
NaCN in 500 ml distilled water, add 2 drops of a
10-percent solution of NaOH, and dilute to 2,000 ml.
The pH of this solution should be about 11.
Hydrobromic acid, 48-percent reagent grade.
Bromine water, saturated solution.
Calcium oxide powder, reagent grade.
Methyl isobutyl ketone (MIBK).
r· Excess air outlet
Santomerse S solution, wetting agent, (
volume;volume in vrater, obtained fr01
Chemical Co., St. Louis, Mo.
APPARATUS
Atomic-absorption ウー・セエイッィュN@ A
Elmer model 303 ato:-nic-absorption spet
meter is used with the following instrum
meters:
Wavelength: 2428 Angstroms.
Slit: 5 (3 millimeters, 20 Angstroms).
Source: Gold hollow cathode. 14 milliampc
rent.
——
Water ャ・カセM r
Figure 1.-Water bath equipped for aeration and constant temperature.
2
Acetylene: 9 pounds per square inch.
Air: 25 pounds per square inch.
Flame adjustments: With the burner lit, adjust the
flow meters to approximately 2.8 on the acety-
Hotplate and magnetic stirrer
Figure 2.-Constant-temperature water bath with magnetic
stirrer.
3
lene and 3.0 on the air while aspirating water.
Minor adjustments may be necessary to obtain
a blue flame when the MIBK is aspirated.
Water bath equipped for aeration and constant temperature
(fig. 1 ). The aeration device constructed in
this laboratory consists of a compressor-vacuum
pump capable of delivering air at one-third cubic
foot per minute under a pressure of 15 po·.mds per
square inch. This amount of air is more than sufficient
for the 20 outlets shown in figure 1. The
compressor supplies a group of air regulator valves
(available from aquarium supply houses) which are
connected to 3/ 16-inch-diameter Pyrex tubes by
Tygon tubing. The valves are used to individually
adjust the air supply to -each sample. A steady air
flow to the tubes is maintained by means of a small
air tank and one or more bleed-off valves.
Constant-temperature water bath with magnetic
stirrer (fig. 2). Teflon-coated セMゥョ」ィ、。ュ・エイ@
magnetic stirring bar.
Pyrex culture tubes, 25 by 200 millimeters, with polyseal
closure screw cap.
CYANIDE ATOMIC-ABSORPTION PROCED’JRE
SAMPLE PREPARATION
Crush sample in a jaw crusher to less than ¥!-inch
size, then split and grind a portion on ceramic plates
set for less than 80 mesh. The sample is no'” sieved;
a large portion of the sample will be less than 200
mesh.
1. Weigh a 2- to 15-g sample and transfer to a porcelain
dish or crucible. Place the crucible in a
cold muffle furnace and heat for 1 hour after the
temperature reaches 600°C. Do not exceed 650°C
because samples of high clay content may tend
to sinter and thus trap the goldandprevent solution
during the cyanide step.
2. Cool. Add 0.25 g CaO powder and mix thcroughly;
transfer sample to a Pyrex culture tube. Add 30
ml of the 0.25-percent NaCN solution to the tube,
place cap on the tube, and shake by hand to wet
the entire sample. Place tube in an automatic
shaker and shake for 15 minutes; let stc.nd overnight.
3. Reshake to break up any cake formed in エィセ@ bottom
of the tube. Wash down sides of the tube with about
10 ml of the NaCN reagent. Add 4 dro’Js of the
0.25-percent Santomerse S wetting agent to the
tube and heat for 30 minutes in the water bath at
80° -85°C while aerating the sample (fig. 1 ).
4. If a 2 -g sample is used, proceed to step 5. If a
sample of more than 2 g is used, centrifuge the
contents of the tube at 2, 000 revolutions per minute
for 5 minutes, then decant the liquid phase
into another culture tube containing a magnetic
stirring bar. Wash the residue left in the first
tube with 10 ml of the 0.25-percent NaCN solution
containing 4 drops of saturated bromine water.
Centrifuge again and combine wash solution with
the original solution.
5. Add 5 drops bromine water and mix. Then, working
under a well-ventilated hood, cautious 1 y add 5
ml of 48-percent HBr. (CAUTION: Hydrogen cyanide
gas is evolved.) Heat in a water bath equipped
for magnetic stirring (fig. 2) for 30 minutes at
90°C,
6. Cool. Remove the stir bar with another magnet held
outside the tube, add 10 ml MIBK, and shake for 1
minute. Centrifuge if step 4 has been omitted.
7. Measure the absorption of the MIBK phase along
with reference standards (see following section,
“Standardization”) on an atomic-absorption spectrophotometer
at a wavelength of2428 Angstroms.
Use a water -saturated so 1 uti on of MIBK as a
reference blank. To obtain a standard calibration
curve, plot the absorption readings against gold
concentration expressed as parts per million in
the organic phase. The gold content of the sample
in parts per million is equal to:
gold in ppm in organic phase X 10 ml
weight of sample in grams
If the sample has an absorbance greater than the
lO ppm standard, a dilution of the organic layer
can be made. The dilution method is illustrated
in the following example: Transfer a 2-ml aliquot
of the organic phase to a tube containing about
30 ml of water and 2 ml of 48-percent HBr. Add
8 ml of MIBK, shake, and measure the absorption.
This dilution factor of 5 will handle a 10-g
sample containing as much as 50 ppm Au.
STANDARDIZATION
Prepare reference standards containing 0.1, 0.2,
0.4, 0.8, 1.0, 2.0, 4.0, 8.0, and 10.0 ppm Au in the
organic phase for working curves by adding appropriate
aliquots of a gold standard solution to a series of
culture tubes. Add 2 ml of 48-percent HBr to each and
adjust the volume to about 40 ml with water. Add 10.0
ml of MIBK and shake for 1 minute to extract the gold
into the organic phase. Measure the absorption of these
standards on the 1X or lOX scales of the atomica
b s or p t i o n spectrophotometer. P e r i o d i c a 11 y,
standards can, and probably should, be checked by
being subjected to the procedure starting with step 2,
However, we have noted no difference in absorption
readings between the standards taken through the entire
procedure and those prepared as described.
FIRE-ASSAY ATOMIC-ABSORPTION PROCEDURE
The fire-assay method involves a fusion of ground
ore with fluxes to form a slag phase and a metallic
lead phase. The lead collects any gold, platinum metals.,
or silver and settles out at the bortmn when poured
into a mold. After cooling., tl’le slag is broken away
from the lead button which is then put into the furnace
in a cupel. The lead melts and is slowly oxidized by
the hot air. The lead oxide wets the semiporous cupel
material, generally bone ash, and is absorbed. The
cupel material is not wet by the molten lead nor by the
noble metals, which remain as a bead after the gradual
absorption of the lead oxide.
4
The procedure for the further treat11ent of the bead
is as follows:
1. Transfer the silver-gold bead obtained by fire assay
to a 50-ml beaker.
2. Add 5 ml of (1:1) HN03, cover, and digest on a
steam bath for 45 minutes. Uncover and evaporate
the solution to about 2.5 ml.
3. Add 6 ml of concentrated HCl, cover, and let stand
20 minutes. Place beaker on steam bath and digest
for 30 minutes, uncover, and evaporate to hard
dryness.
4. Add 7 ml of concentrated HCl, cover, and digest on
the steam bath until all salts are in solution. Uncover,
add about 15 ml of water, and digest an
additional 30 minutes.
5. Cool. Quantitatively transfer the solution to a Pyrex
culture tube by washing beaker with about 10 ml
of water. Add 2 ml of 48-percerr: HBr, mix, add
10 ml of MIBK, and shake tube for 1 minute.
6. Determine the absorbance of the organic phase along
with reference standards prepared in the same
manner as described in the cyc.nide procedure.
We routinely add 2 milligramsofsilverinthe fusion
step. Precut lengths of lead rod, eac’1 containing this
amount of silver, are obtainable commercially under
the trade name Inquarts. Silver is important in this
procedure for three reasons: (1) It has a protective
effect and thus reduces gold losses c’uring the cupellation
step (Bugbee, 1940, p. 107), (2) it allows trace
amounts of gold to be easily transferred from the
cupel to a suitable container for analysis, and (3) the
addition of silver provides a silver-gold bead that is
easily dissolved, The ratio of silver to gold should be
greater than 3:1 to allow dissolutioncfthebead; beads
having lower ratios of silver to gold may be difficult
to dissolve, even in aqua regia. For ウセューャ・@ that contain
sufficient silver, such addition is unnecessary;
conversely, for samples than contain considerable
gold, the addition of more than 2 rrilligrams silver
may be necessary to attain this 3:1 ratio. The silvergold
bead is routinely weighed to allow an estimate
or approximate check of the combined gold and silver
content of the ore,
For a full discussion of the fire-assay method the
reader is referred to Bugbee ( 1940) or Beamish ( 1966 ).
Table 1.-Comparison of gold values obtained by 。エッュゥ」セ「ウイーョ@ methods and by the conventional gravimetric
ヲゥイ・セ。ウケ@ method
[Gravimetric fire-assay by 0, M. Parker and L, B. Riley, Atomic-absorption determinations by
Claude Huffman, Jr., J.D. Mensik, and L. B. Riley]
Gold (ppm)
Sample Atomic absorption
Sample description
Gravimetric and locality
fire assay Fire assay Cyanide
96130 _______________
4,8 4.8
0223180 ———— 10,3 11
Dll9999 ———— 13,7 16
D100874 ———— 15.8 18
Dll9971 ———— 16.5 15
D119956 ———— 26.4 27
RESULTS
The accuracy of the two proposed atomic-absorption
methods was first tested by analyzing a suite of six
selected samples previously an a 1 y zed by the conventional
gravimetric fire-assay method. The results
(table l) show good agreement between the two atomic
absorption methods and the gravimetric fire-assay
method. These samples were selected because they
represent a wide variety of rock and ore types as
indicated by semiquantitative spectrographic analysis.
For example, sample Dll9956 contains 1.5 percent As,
7 percent Fe, and 0.07 percent Zn; sample Dl19971
contains 710 percent As, lOpercent Fe, and 0.03 percent
Sb; sample Dll9999 contains> lO percent Fe, 5
percent Cu, and 0.2 percent Mn. About 90 percent of
sample Dl00874 is estimated to be pyrite.
The cyanide atomic-absorption method has been
thoroughly tested by routinely analyzing several hundred
samples of varying composition. All samples that
contained at least 0. 5 ppm Au were reanalyzed by the
fire-assay atomic-absorption method; in addition,
many randomly selected samples containing less than
0.5 ppm Au were checked. Table 2 presents some of
the data on 20 samples from Alaska and Nevada areas.
The cyanide and fire-assay methods show concordant
results.
SAMPLE DISSOLUTION, CYANIDE PROCEDURE
Simmons (1965) observed that low results were obtained
on some samples if the ignition step was omitted
prior to the cyanide treatment. We have confirmed his
observation and routinely roast all samples.
The cyanide process has been used commercially for
many years. The commercial processes are based on
5
4.3 Sulfide ore, White Pine mine,
Jefferson County, Mont.
10 Channel sample, Gold Anchor
mine, Clear Creek County, Colo.
15 Pyrite, Billy Goat mine, Okanogan
County, Wash.
14 Quartz pyrite ore, Phoenix mine,
Clear Creek County, Colo.
16 Quartz vein, Okanogan County,
Wash.
25 Mineralized quartzite, Okanogan
County, Wash.
the following chemical reaction:
4Au+8NaCN+2Hz0+02 -‘)oo4AuCN·NaCN+4NaOH.
The amount of gold which can be dissolved in sodium
cyanide solution depends on the amount of oxyr-en present
(Jacobson, 1949, p. 554). The aeraticndevice
described in this paper supplies the oxygen for the
preceding reaction and furnishes agitation to a large
sample. It is desirable to keep the sample ir a single
tube during the entire analysis. In the ir.itial stage of
this investigation, 2 -g samples were aeratec, conveniently
by means of magnetic stirring. However, when
we attempted to increase our sample size to 15 g we
found that the magnetic stirrer would not sufficiently
agitate the larger sample. The aeration device provides
sufficient agitation of a 15-g sample.
EXTRACTION OF GOLD WITH METHYL ISOBUTYL
KETONE
Methyl isobutyl ketone (MIBK) has been used extensively
in atomic-absorption work to preconcentrate
certain elements that show insufficient ウ・ョセゥエカケ@ in
aqueous solution. Many authors favor this reagent
because it has excellent flame characteristics and it
is only slightly soluble in water ( 1. 9 g per 100 ml).
We wish to emphasize that the gold is in the monovalent
state after the cyanide treatment and will not be
extracted quantitatively into the MIBK unlesE’ oxidized
to the trivalent state. In thi’s method the dicyanogold
(I) ion is oxidized to the bromaurate (III) ion by the
addition of bromine water and hydrobromic acid. The
reaction is as follows:
n。aオHcnIコKRhbイKbRセbSKRhcnKnN@
The bromaurate (III) ion is then extracted quan’:itatively
into MIBK prior to estimation by atomic ab”‘orption.
Table 2.-Comparison of gold values obtained by fireassay
atomic-absorption and cyanide atomic-absorption
methods
[Analyses by Claude Huffman, Jr., J. D.
Mensik, and L. B. Riley1
Gold (ppm)
Sample
Sample
Fire Cya- description
assay nide
D124654 __ 0,05 0.05 Altered rock.
ABH775 __ . 09 .07 Altered diorite .
D124913 __ .09 . 1 Altered zone .
ABH780 __ .4 . 3 Iron -stained quartz .
D124670 __ . 7 ,8 Altered zone .
D124667 __ . 9 ,8 Quartz vein .
D124665 __ 1 1 Altered zone.
Dl24943 __ 2 2 Massive sulfide
vein.
ABH923 __ 2 2 Stained silicified
quartzite.
D124916 __ 3 3 Sulfide replacement
lenses.
ABH926 __ 4 4 Drusy quartz and
iron oxide,
D12493l__ 7 5 Altered zone and
quartz vein,
ABH871 __ 9 8 Gossan zone in
quartzite,
D124676 __ 12 13 Quartz vein and
altered zone,
0124675 __ 17 16 Quartz vein.
D124920 __ 27 27 Sulfide -bearing
quartz -calcite
vein.
ABH854 __ 44 40 Quartz, calcite,
sulfides.
D124679 __ 52 58 Altered zone.
ABH832 __ 85 73 Iron-stained quartz.
D124681 __ 350 330 Altered zone.
Test solutions containing 1, 10, 50, and 100 micrograms
of monovalent gold in 40 ml of cyanide solution
were oxidized by adding 5 drops of saturated bromine
water and 5 ml of 48-percent HBr, heated to expel
hydrogen cyanide gas, cooled, and extracted with 10 ml
of MIBK. All these test solutions showed by atomicabsorption
spectrophotometry a recovery of greater
than 95 percent in the organic phase.
Recently. Strelow, Feast, Mathews, and Van Zyl
セ@ 1966) 」セョヲQイュ・、@ our observation that monovalent gold
m cyamde solution is not directly extractable with
MIBK; they used KMn04 in HCl solution to oxidize the
gold to the trivalent state.
Atomic-absorption methods for the determination of
gold in solution are reasonably free from interferences.
No effect from diverse ions has been noted in
6
this work. In the fire-assay bead method, silver is
partially extracted into the MIBK, but haS” no effect on
the gold determination. Strelow, Feast, Mathews, and
Van Zyl (1966) also observed no interferences; they
stated that Tl(lll), Cr(Vl), Mo(Vl), As(V), セijHv@ ), Sn(IV),
Se(l V ), Te(IV), and Fe(III) are partially extracted but
cause no interference with the determination of gold
by atomic absorption.
CONCLUSIONS
The fire-assay separation combined with atomic
absorption has two advantages over the classical fireassay
method: ( 1) It eliminates the tedious and critical
steps of transferring, parting, annealing, and then
weighing trace amounts of gold and (2) it provides
greater accuracy and precision in low and fractional
parts-per-million range.
We employ the more rapid cyanide method for scanning
large numbers of samples for their gold content.
Samples that show a gold content of greater than 1 ppm
are then reanalyzed by the fire-assay atcmic-absorption
method to confirm the results.
REFERENCES
Beamish, F. E., 1966, The analytical chemistry of the
noble metals in International series of monographs
in an a 1 y tic a 1 chemistry [1st ed. J: New York,
Pergamon Press, v. 24. 609 p.
Bugbee, E. E., 1940, A textbook of fire assaying [3d
ed.]: New York, John Wiley &Sonrc,Inc.,314p.
Jacobson, C. A., ed., 1949, Encyclopedia of chemical
reactions: New York, Reinhold Put. Corp., v. 3,
800 p.
Lakin, H. W., and Nakagawa, H. M., 1965, A spectrophotometric
method for the de t e r m i nat ion of
traces of gold in geologic materials, in Geological
Survey research 1965: U.S. Geol. Survey Prof.
Paper 525-C, p. Cl68-C171.
Olson, A. M., 196.5, Gold assay by atomic absorption
spectrophotometry: Perkin-Elmer Atomic Absorption
Newsletter, v. 4, no. 5, p. 278-280.
Simmons, E. C., 1965, Gold assay by atomic absorption
spectrophotometry: Perkin- Elmer Atomic
Absorption Newsletter, v. 4, no. 2, p. 281-287.
Strelow, E. C .. Feast, P. M., Mathews, C. J. C., and
Van Zyl, C. R., 1966, Determination of gold in
cyanide waste solutions by solvent extraction and
atomic a b so r p t ion spectrophotometry: Anal.
Chemistry, v. 38, p. 115-117.
Tindall, F. M., 1965, Silver and gold assay by atomic
absorption spectrophotometry: Perkin-Elmer
Atomic Absorption Newsletter, v. 4, no. 9, p. 339-
340.
Walsh, Ao, 1955, The application of atomic absorption
spectra to chemical analysis: Spectrochim. Acta,
v. 7, p. 108-117.